Biomedical Test System
Defibrillator Module
9
9-9
Out-of-Limits Reading
When a reading is out of limits, a # sign appears next to the reading with the
statement:
Out of limits
. Both the high and low limits will appear. Whenever
this occurs, you can repeat the test or continue with the remaining steps.
Press
F4
to repeat the test or
F5
to continue to the next step.
In order to flag a defib reading as out of limit, the reading must exceed the
programmed test limit percentages (
±
) and differ from the expected step
energy value by more than one (1.0) joule. This out-of-limit determination
technique allows for a wider range of acceptable reading at the lower energy
settings which is typical of defibrillator specifications. For example, a
defibrillator can have a
±
50 percent tolerance at the low 2.0 joule (1.0-3.0)
which is covered by the
±
1.0 joule test limit and could have failed if only a
±
10 percent (1.8-2.2) test limit was used.
After the low range steps execute, you receive a prompt to set the Impulse to
the high range:
You can now execute the high range energy tests. These steps occur in the
same order as described above.
Note
You can skip an energy step by pressing the down arrow key on
the medTester. Whenever you do this, the energy reading line
displays the nominal value for the step followed by an
X
. Here's
an example:
ENERGY = 100.X J.
Set Impulse to HIGH range
READY
F5
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...