Biomedical Test System
SPO2 Module
15
15-21
For the Index 2XL:
SPO2 Autosequence Number: A74
Autosequence Name: GENERC
Manufacturer: Generic–Not Specified
Settling Time: 25 Beats
Heart Rate Limit:
±
4 BPM (beats per minute)
Step Number
SPO2
(in percent
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Base TLC
1 99
±
4
250 6.0 4000
2 96
±
4
230 6.0 4000
3 92
±
4
205 6.0 4000
4 88
±
5
180 6.0 4000
5 78
±
5
155 6.0 4000
6 75
±
5
130 6.0 4000
7 65
±
6
105 6.0 4000
8 55
±
6
80 6.0 4000
9 45
±
6
55 6.0 4000
10 35
±
6
30 6.0 4000
11 90
±
4
95 6.0 4000
12 88
±
5
45 6.0 4000
13 85
±
5
130 2.0 4000
14 92
±
4
95 3.0 4000
15 93
±
4
90 4.0 4000
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 6.0%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
No probe test included
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...