Biomedical Test System
Installation
2
2-3
medTester 5000C Power-Up Condition Messages
Message
Condition Indicated
LOW BATTERY, SERVICE REQUIRED!
Internal battery (used for memory and clock
backup) level is low.
LOW LINE VOLTAGE!
Hot to neutral is less than (<) 90 V.
HIGH LINE VOLTAGE!
Hot to neutral is greater than (>) 135 V.
OPEN GROUND OR ISOLATED POWER SYSTEM! Hot to ground differs from neutral to ground by
more than 50 V.
REVERSE POLARITY OUTLET!
Neutral to ground is greater than (>) 70 V and hot
to ground less than (<) 35 V.
Initialization
You can initialize your medTester with two different local keyboard
commands which you can type at the main menu.
•
INIT +
Enter—
Initializes autosequences and custom features with
YES or NO prompts for each section. This initialization command
does not erase test records or checklists from memory
•
I
Enter—
Initializes the entire instrument with no
exceptions. All test records and checklists are erased from memory
with this option.
Date and Time Setup
Date uses the
month/date/year
format, and time uses the
hour:minute:second
format. Note that time is entered in the 24-hour format. To reset the date and
time:
1.
Enter menu commands:
UTIL
F5
CLOCK
F4
See note below.
2.
Enter the date and time directly from the keyboard, or…
Use the keyboard left and right arrow keys to move to the desired
month/date/year or hour/minute/second data entry point.
3.
Enter the desired data from the keyboard, or use the up/down
keys to increment or decrement the data.
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...