medTester 5000C
Operators Manual
12-8
UNIT of MEASUREMENT ABBREVIATIONS
Abbreviation
Unit of Measurement
V RMS
Volts RMS
A RMS
Amperes RMS
µA RMS
Microamperes RMS
µA DC
Microamperes DC
OHMS Ohms
RPM
Revolutions per minute
DEG F
Degrees Fahrenheit
DEG C
Degrees Centigrade
% Percent
Error Responses
If you issue an incorrect or nonexistent remote command, the medTester
responds with an error statement in this format:
ERR=
xx. Here is a list of
possible error responses and explanations.
ERROR RESPONSES
Error Statement
Explanation
ERR=01,UNKNOWN COMMAND
The received command was not
recognized.
ERR=02,ILLEGAL COMMAND
The received command was not
allowed for the current instrument
status.
ERR=03,ILLEGAL PARAMETER
A parameter sent with a command
was not the right type, was out of
range, or was not understood, or
not the right number of parameters
were sent.
ERR=04,MODULE NOT INSTALLED
Can not execute the command
because the required instrument
module is not installed.
ERR=20,CURRENT SOURCE NOT
CONNECTED
For a resistance measurement
command, the current source
connection has not been made.
ERR=21,MEMORY FULL
For a file down transfer command,
the medTester memory is full and
can not accept anymore data to
fulfill the transfer.
ERR=22,GFCI
A ground fault circuit interruption
was triggered caused by a fault in
the EUT plugged into the test
receptacle. This error is not a
response to a command. It
appears whenever a ground fault
occurs.
ERR=23,WRONG FILE
For a file transfer command where
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...