medTester 5000C
Operators Manual
4-10
External Resistance
See Figure 4-10,
External Resistance
.
1.
Connect the dual banana-plug end of one Kelvin cable to the two
red
external meter posts.
2.
Connect the dual banana-plug end of the other Kelvin cable to the
two
black
external meter posts.
3.
Connect the alligator-clip ends of the two Kelvin cables to
opposite ends of the resistance being measured.
4.
Enter menu commands:
OHMS
F4
MANUAL
F2
EXT
F2
The
CURRENT SOURCE
lamp lights up to indicate that the
connection is made. The display reads the resistance (in OHMS).
Isolated Power and Ground Fault Test
See Figure 4-11,
Isolated Power System and Ground Fault Test.
This test requires the Fluke Biomedical Model 202A
Isolated Power Test Module
plugged into the test receptacle of the medTester 5000C. The Model 202A
introduces faults on isolated power systems and grounded systems.
Refer to Model 202A Operating and Service Manual for details on ground
fault and isolated power testing.
Enter menu commands:
MANUAL
F2
ISOPWR
F5
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...