7-8
Section 7 – Remote Operation
STATUS
(Continued)
Each register set consists of three individual registers, as depicted in
Figure 7-3
.
The CONDITION register is a real time, read-only register that is constantly
updated to reflect current operating conditions.
The EVENT register is fed by the CONDITION register, but operates as a latch.
Whenever any bit in the CONDITION register goes to “1”, a corresponding “1” is
latched into the EVENT register and remains that way until cleared by a specific
command.
The ENABLE register is a mask register that is used to generate the single status
bit in the STATUS BYTE. Setting any bit in the ENABLE register to “1” will allow
a corresponding “1” in the EVENT register to set the summary bit in the STATUS
BYTE.
An example of their usage appears at the end of this section.
Figure 7-3 Condition, Event, and Enable Registers
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...