6-4
Section 6 – Flux Density Measurement
RANGE
SELECTION
The instrument is capable of providing flux density measurements on one of
four fixed ranges, or it can be programmed to automatically select the best
range for the present flux density being measured. The available ranges are
listed in
Section 2 – Specifications
. The ranges advance in decade steps,
the magnitudes available depend on the probe type. The lowest range offers
the best resolution while the highest range allows higher flux density levels to
be measured.
Setting the Range
Pressing the RANGE key will advance the range setting. The four ranges
are followed by the autorange selection.
Autorange
In the autorange mode, the range is advanced if the reading reaches 90% full
scale of the present range. The range is lowered if the present reading falls
below 8% full scale of the present range.
NOTE: When the word “
Auto
” does not appear on the “measurement
mode indicator” line, the channel is in manual ranging mode.
Overrange Condition
An overrange condition will occur when the following is true:
The instrument is in manual range mode or is in autorange mode
at the highest range; and the measured flux density is 110% of the
present full scale range.
When an overrange condition occurs the instrument will display the message
“
OverRange”
in place of the flux density reading.
The next highest range
should be selected. If already on the highest range, then the flux density is
too great to be measured with this instrument and its probes.
Note:
For dc mode operation, the polarity of the flux density that caused
the overrange condition will be displayed to the left of the
“
OverRange”
message.
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...