![F.W. Bell 7030 Скачать руководство пользователя страница 105](http://html.mh-extra.com/html/f-w-bell/7030/7030_instruction-manual_535140105.webp)
F.W. BELL Model 7030 Gauss/Tesla Meter Instruction Manual
Section 7 – Remote Operation
7-31
SCPI COMMANDS-
CLASSIFIER (LIMIT)
COMMANDS
(Continued)
:CALCulate#:LIMit:LOWer:DATa <NRf>
:CALCulate#:LIMit:UPPer:DATa <NRf>
<NRf> is a signed decimal number that sets the lower and upper
classifier limits. <NRf> must be between 0.0000000000 and
±
29999999. The meaning of the limit depends upon the currently
selected unit of measure (gauss, tesla, etc.). For instance the value
1.0203 could mean 1.0203 G or 1.0203 T.
If a lower limit is specified that is arithmetically larger than the upper
limit, the two limits will be reversed internally. The sign of the limit
has no meaning in the ac measurement mode.
:CALCulate#:LIMit:LOWer:DATa?
:CALCulate#:LIMit:UPPer:DATa?
The requested classifier limit is placed in the output queue. The limit
will be between 0.0000000000 and
±
29999999. The meaning of the
limit depends upon the currently selected unit of measure (gauss,
tesla, etc.). For instance the value 1.0203 could mean 1.0203 G or
1.0203 T.
:CALCulate#:LIMit:STATe <b>
The classifier function is turned on when <b> is “1” or “ON”, or turned
off when <b> is “0“ or “OFF”.
:CALCulate#:LIMit:STATe?
The state of the classifier function (“OFF” or “ON”) is placed in the
output queue.
:CALCulate#:LIMit:FAIL?
The pass/fail status of the classifier test is placed in the output queue.
A “0” indicates that the measured value falls between the lower and
upper limit and is considered a “pass” situation. A “1” indicates that
the measured value either falls below the lower limit or above the
higher limit.
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...