F.W. BELL Model 7030 Gauss/Tesla Meter Instruction Manual
6-6
Section 6 – Flux Density Measurement
AC MODE
OPERATION
(Continued)
Frequency Compensation in ac Mode
To obtain accurate ac measurements using frequency compensation, the
ac field must have a frequency greater than 10 Hz, and depending on the
selected range, a minimum magnitude according to
Table 6-2.
Table 6-2 Minimum Magnitudes for Rated ac Accuracy
Range
Minimum Magnitude
Low Field Probe
Mid Field Probe
High Field Probe
300.000 mG (30.0000
µ
T)
30.0000 G (3.00000 mT)
300.000 G (30.0000 mT)
20% Full Scale
3.00000 G (300.000
µ
T)
300.000 G (30.0000 mT)
3.00000 kG (300.000 mT
6% Full Scale
N/A
3.00000 kG (300.000 mT)
30.0000 kG (3.00000 T)
4% Full Scale
N/A
30.0000 kG (3.00000 T)
300.000 kG (30.0000 T)
2% Full Scale
If the measured magnetic field is less than the “minimum magnitude”
shown on
Table 6-2
, fourth column; the frequency of the magnetic
field cannot be measured. In that case, the frequency indicator will
display dashes. See
Figure 6-4
.
Figure 6-4 Indeterminate Frequency Indicator
------Hz f5k
Note: Although the flux density reading will be displayed when the
frequency is indeterminate, it is not specified at rated
accuracy.
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...