![F.W. Bell 7030 Скачать руководство пользователя страница 103](http://html.mh-extra.com/html/f-w-bell/7030/7030_instruction-manual_535140103.webp)
F.W. BELL Model 7030 Gauss/Tesla Meter Instruction Manual
Section 7 – Remote Operation
7-29
SCPI COMMANDS-
RANGE COMMANDS
(Continued)
:SENSe#:FLUX:RANGe?
Places an ASCII string in the output queue representing the present
ac/dc setting of the channel, the present range and the state of the
AUTO range feature.
The string will start with the phrase “AC” or “DC, followed by a comma(,).
The next single digit will be the range setting, as shown in
Table 7-6,
and will be in the range of 1 to 4, followed by a comma(,). Next will be
the phrase “OFF” or “ON, indicating the state of the AUTO range
feature.
For example the string DC,2,OFF indicates that the channel is in dc
mode on a fixed range setting of 2.
SCPI COMMANDS-
FILTER COMMANDS
These commands control the hardware low-pass filters.
Since these commands are channel-specific the “#” in each command
should be replaced with the channel number 1,2 or 3. If the channel
number is omitted then Channel-1 is assumed.
:SENSe#:FLUX:FILTer:LPASs:FREQuency <NRf>
There are two choices for the cutoff frequency for the low pass filter; 500
Hz or 5 kHz. These are two-pole filters with an attenuation of –12 db per
octave. <NRf> specifies the cutoff frequency in hertz, either 500 or
5000.
:SENSe#:FLUX:FILTer:LPASs:FREQuency?
Places the low-pass filter’s cutoff frequency in hertz in the output queue,
either 500 or 5000.
:SENSe#:FLUX:FILTer:LPASs:STATe <b>|AUTo
The low-pass filter is turned off when <b> is “0” or “OFF”, on when <b>
is “1” or “ON”, or placed in AUTO mode then <b> is “AUTo”.
When the low pass filter is turned off, the instrument has an inherent
cutoff of 50kHz.
:SENSe#:FLUX:FILTer:LPASs:STATe?
The state of the low-pass filter (”OFF”, “ON” or “AUTO”) is placed in the
output queue.
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...