![F.W. Bell 7030 Скачать руководство пользователя страница 106](http://html.mh-extra.com/html/f-w-bell/7030/7030_instruction-manual_535140106.webp)
F.W. BELL Model 7030 Gauss/Tesla Meter Instruction Manual
7-32
Section 7 – Remote Operation
SCPI COMMANDS-
ZEROING
COMMANDS
These commands control zeroing of the Hall probe and the channel to
which it is attached.
Since these commands are channel-specific the “#” in each command
should be replaced with the channel number 1,2 or 3. If the channel
number is omitted then Channel-1 is assumed. As a special case a
channel number of 4 will be accepted and will cause all active channels
to be zeroed simultaneously.
:CALibration#:ZERo:HSENsor:INITiate
This command initiates the same zeroing process as the front-panel
ZERO push button. The Hall probe should be placed in a stable, low-
flux environment such as a zero flux chamber prior to issuing this
command. While zeroing is in progress the “ZERO” status bit in the
OPERATION CONDITION REGISTER will be set.
This command does not generate a response.
:CALibration#:ZERo:HSENsor:INITiate?
Same as the “:CALibration#:ZERo:HSENsor:INITiate” command, but a
response is returned upon completion of the zeroing process. A “0”
indicates successful completion, otherwise a non-zero number is
returned indicating a failure.
SCPI COMMANDS-
RELATIVE OFFSET
COMMANDS
These commands control the RELATIVE function. The RELATIVE
function subtracts a fixed flux density value from all measured values.
The user can define the offset or it can be generated using the present
value of flux density as measured by the Hall probe.
Since these commands are channel-specific the “#” in each command
should be replaced with the channel number 1,2 or 3. If the channel
number is omitted then Channel-1 is assumed.
There are some limitations when using the RELATIVE function. Refer to
Section 6 - Flux Density Measurement
for details.
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...