F.W. BELL Model 7030 Gauss/Tesla Meter Instruction Manual
Section 7 – Remote Operation
7-35
SCPI COMMANDS-
ANALOG OUTPUT
COMMANDS
(Continued)
Some data acquisition systems may require different analog input
ranges. For example, assume such a system cannot accept signals
greater than
±
2.0 volts. In this case you can specify 2.0V as the full-
scale range of the analog output (<NRf> = 2.0).
The relationship between the output voltage and the flux density reading
is:
Full scale output voltage / full scale flux density range
For example, assume the full scale output voltage has been set to 2.0
volts and the present range setting is 300 mT. The relationship will be:
2.0 volts / 300 mT = 0.0067 volts / mT
Thus, when the flux density reaches the maximum for the present range,
the analog output voltage will reach a maximum of
±
2.0 volts.
:OUTPut#:ATTenuation1?
The full scale voltage setting of the uncorrected analog output (0.0 to
10.0) is placed in the output queue.
:OUTPut#:COUPling2 AC|DC
When the channel is configured for alternating (ac) flux density
measurement you can select one of two signals to appear at the
corrected analog output. When “AC” is specified the signal will be the
real-time waveform as measured by the Hall probe, within the low
frequency limits of the corrected output. When “DC” is specified the
output signal will be a dc level representing the true rms value of the low
frequency waveform.
When the channel is configured for static (dc) flux density measurement
you can select one of two signals to appear at the corrected analog
output. When “AC” is specified the signal will be the real-time, low
frequency waveform as measured by the Hall probe, within the low
frequency limits of the corrected output. When “DC” is specified the
output signal will be a dc level representing the true rms value of the low
frequency waveform.
NOTE: This command does not affect the vector summation output. If “4”
is specified as a channel number (#), an error will be reported.
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...