![F.W. Bell 7030 Скачать руководство пользователя страница 102](http://html.mh-extra.com/html/f-w-bell/7030/7030_instruction-manual_535140102.webp)
F.W. BELL Model 7030 Gauss/Tesla Meter Instruction Manual
7-28
Section 7 – Remote Operation
SCPI COMMANDS-
RANGE COMMANDS
These commands select either a fixed range or AUTO range for each
channel, and specify if the measured flux density will be static (dc) or
alternating (ac).
Since these commands are channel-specific the “#” in each command
should be replaced with the channel number 1,2 or 3. If the channel
number is omitted then Channel-1 is assumed.
:SENSe#:FLUX:AC
:SENSe#:FLUX:DC
Specifies that the measured flux density will be static (dc) or alternating
(ac). These commands do not alter the present range setting.
:SENSe#:FLUX:RANGe:FIXed <n>
:SENSe#:FLUX:DC:RANGe:FIXed <n>
:SENSe#:FLUX:AC:RANGe:FIXed <n>
Specifies that the measured flux density will be static (dc) or alternating
(ac). If the “AC” or “DC” key word is omitted, then static (dc) operation is
selected. Also specifies one of four fixed operating ranges for the
instrument, as defined in the following table:
Table 7-6 Operating Ranges with Various Probes
<n>
Maximum Flux Density
Using Low Field Probe
Maximum Flux Density
Using Mid Field Probe
Maximum Flux Density
Using High Field Probe
1
300 mG (30
µ
T)
30 G (3 mT)
300 G (30 mT)
2
3 G (300
µ
T)
300 G (30 mT)
3 kG (300 mT)
3
30 G (3 mT)
3 kG (300 mT)
30 kG (3 T)
4
300 G (30 mT)
30 kG (3 T)
300 kG (30 T)
:SENSe#:FLUX:RANGe:AUTo <b>
:SENSe#:FLUX:DC:RANGe:AUTo <b>
:SENSe#:FLUX:AC:RANGe:AUTo <b>
Specifies that the measured flux density will be static (dc) or alternating
(ac). If the “AC” or “DC” key word is omitted, then static (dc) operation is
selected.
AUTO ranging is turned off when <b> is “0” or “OFF”, or on when <b> is
“1” or “ON”. When AUTO range is turned on the instrument determines
the range that will offer the best resolution. When AUTOrange is turned
off the range will be fixed to the last range that was in use when the
command was issued.
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...