F.W. BELL Model 7030 Gauss/Tesla Meter Instruction Manual
7-34
Section 7 – Remote Operation
SCPI COMMANDS-
ANALOG OUTPUT
COMMANDS
Each channel provides an uncorrected analog output signal that originates
from the amplifier section of the instrument. This signal is not corrected for
errors in the Hall probe or instrument but is capable of high frequency
operation. Also provided is a corrected output signal that is generated
digitally. This signal is corrected for most errors in the Hall probe and
instrument, but is only capable of low-frequency reproduction. One
additional output represents the vector summation result of all three
channels. This is also considered to be a corrected output.
You can define the type of signal that appears at each output and the scale
of that signal, from the ANALOG OUT menu or by issuing these commands
Since these commands are channel-specific the “#” in each command
should be replaced with the channel number 1,2 or 3. If the channel
number is omitted then Channel-1 is assumed. In some commands a
channel number of 4 will be accepted to specify the vector summation
output.
:OUTPut#:COUPling1 AC|DC
When the channel is configured for alternating (ac) flux density
measurement you can select one of two signals to appear at the
uncorrected analog output. When “AC” is specified the signal will be the
real-time waveform as measured by the Hall probe. When “DC” is
specified the output signal will be a dc level representing the true rms value
of the waveform.
When the channel is configured for static (dc) flux density measurement
you can select one of two signals to appear at the uncorrected analog
output. When “AC” is specified the signal will be the real-time, low
frequency waveform as measured by the Hall probe, from 0 – 400 Hz.
When “DC” is specified the output signal will be a dc level representing the
true rms value of the low frequency waveform.
:OUTPut#:COUPling1?
The type of signal selected for the uncorrected analog output (“AC” or
“DC”) is placed in the output queue.
:OUTPut#:ATTenuation1 <NRf>
The full scale output voltage of the uncorrected analog output can be
specified to any level between 0 and 10 volts in 0.1 volt steps. Normally a
full-scale level of 3.0 volts is used because it directly corresponds to the
full-scale ranges of the instrument. For instance, if on the 300 mT range, a
reading of -189.3 mT will result in an analog output of -1.893 volts. If on
the 3 kG range, a reading of +2.6 kG will result in an analog output of +2.6
volts.
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...