SCPI COMMANDS-
RELATIVE OFFSET
COMMANDS
(Continued)
:INPut#:OFFSet <NRf>
This command defines a fixed relative offset value, which will be
subtracted from all measured values. The relative offset value will only
be used when the RELATIVE function is enabled. See the
“:INPut#:OFFSet:STATe” command for details.
<NRf> is a signed decimal number between 0.0000000000 and
±
29999999. The meaning of the offset depends upon the currently
selected unit of measure (gauss, tesla, etc.). For instance the value
1.0203 could mean 1.0203 G or 1.0203 T.
:INPut#:OFFSet?
The relative offset value is placed in the output queue and will be
between 0.0000000000 and
±
29999999. The meaning of the offset
depends upon the currently selected unit of measure (gauss, tesla,
etc.). For instance the value 1.0203 could mean 1.0203 G or 1.0203 T.
:INPut#:OFFSet:STATe <b>|ONCe
When <b> is “0” or “OFF” the RELATIVE function is turned off. The
present flux density readings will not be altered. The previously defined
relative offset value will be preserved.
When <b> is “1” or “ON” the RELATIVE function will be enabled. A
fixed relative offset value will be subtracted from all flux density
readings. This value could have been previously defined using the
“:INPut#:OFFSet <NRf>” command, manually entered using the
RELATIVE menu, or generated automatically using the probe as the
input source.
When <b> is “ONCe” the automatic RELATIVE operation is started.
This command initiates the same process as the front-panel RELATIVE
push button. The Hall probe should be placed in a stable environment
prior to issuing this command. The value presently measured by the
probe will become the new relative offset value. This will overwrite any
value previously defined using the “:INPut#:OFFSet <NRf>” command
or entered manually from the RELATIVE menu. When the automatic
RELATIVE operation is complete the RELATIVE function will be
enabled.
:INPut#:OFFSet:STATe?
The state of the RELATIVE function (“OFF” or “ON”) is placed in the
output queue.
Содержание 7030
Страница 1: ...7030 GAUSS TESLA METER Instruction Manual ...
Страница 2: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual THIS SIDE BLANK Inside of Front Cover ...
Страница 24: ......
Страница 25: ...THIS SIDE BLANK Rear of page 2 7 ...
Страница 26: ......
Страница 38: ...F W BELL Model 7030 Gauss Tesla Meter Instruction Manual 4 8 Section 4 Setup This Page Intentionally Blank ...
Страница 122: ...THIS SIDE BLANK Back B 3 ...
Страница 124: ...THIS SIDE BLANK Back of Warranty ...
Страница 125: ...THIS SIDE BLANK Inside of Rear Cover ...
Страница 126: ...4607 SE International Way Milwaukie OR 97222 Phone 503 659 5999 www oeco com ...