-43-
5-8 Operations of related units connected via NVC communication
5-8-1. Evaporator
When you connect a related units with an
NVC communication cable and turn its power
ON, communication will start.
LCD display shows measured values and set
values.
(
Control mode select screen, setting screen,
graph screen, solvent registration screen,
database screen)
Mode
MANUAL
Standby
00:00:00
Display
1013
hPa
Set Press
hPa
200
Rotation
180
rpm
Display example when model N-1210B
is connected
Rotation display
Lower:Set value
Mode
MANUAL
Standby
00:00:00
Display
1013
hPa
Set Press
hPa
200
Rotation
0
rpm
180
Display example when model N-
1300 is connected
Rotation display
Upper:Measured value
Lower:Set value
Leak
Menu
Run
Stop
Power
Mode
MANUAL
Standby
00:00:00
Display
1013
hPa
Set Press
hPa
200
Leak
Menu
Run
Stop
Power
Rotation 0rpm
180
Vapor 27.6
℃
Bath 40.0
℃
40.0
Circulater
5.0
℃
5
Rotation display
Upper:Measured value
Lower:Set value
Bath temp display
Upper:Measured value
Lower:Set value
Cooling temp display
Upper:Measured value
Lower:Set value
(1) Model N-1300 evaporator
Measured rotations and settings are displayed.
When operation of NVC is started, N-1300 starts
rotation and when operation of NVC is stopped, rotation
of N-1300 stops.
When rotation of N-1300 is started, operation of NVC
starts and when operation of N-1300 is stopped,
operation of NVC stops when rotation of the flask is
stopped.
When an alarm occurs in any of other units connected
with the NVC communication cable, “(12)System
Alarm” is triggered and rotation of N-1300 is stopped.
(2) Model N-1210B evaporator
Set number of rotations is displayed.
When operation of NVC is started, rotation of N-1210B
starts and when operation of NVC is stopped, N-1210B
will be jacked up automatically after its rotation has
stopped.
When rotation of N-1210B is started, operation of NVC
starts and when operation of N-1210B is stopped,
operation of NVC is stopped and jacked up
automatically after rotation of the flask has stopped.
When an alarm occurs in any of other units connected
with the NVC communication cable, “(12)System
Alarm” is triggered and N-1210B will be jacked up
automatically after its rotation has stopped.
(3) Model N-2110 evaporator
Set number of rotations, measured value of the bath and
the set value are displayed.
When operation of NVC is started, rotation of N-2110
starts and when operation of NVC is stopped, N-2110
will be jacked up automatically after its rotation has
stopped.
When rotation of N-2110 is started, operation of NVC
starts and when rotation of N-2110 is stopped, operation
of NVC is stopped and will be jacked up after rotation
of the flask has stopped.
When an alarm occurs in any of other units connected
with the NVC communication cable, “(12)System
Alarm” is triggered and temperature control of N-2110
bath is stopped and will be jacked up automatically after
its rotation has stopped.
Mode
MANUAL
Standby
00:00:00
Display
1013
hPa
Set Press
hPa
200
Bath 40.0
℃
40
Bath temp display
Upper:Measured value
Lower:Set value
Rotation display
Lower: Set value
Rotation
160
rpm
Display example when model N-2110 is
connected