(3) Setting of operation stop conditions (Auto Stop)
Setting range
:
Off, 1
~
999min(unit:1)
Temp (When the vapor temperature sensor is
connected)
Initial value
:
Off
The unit will stop automatically when the set time has
elapsed since the start of operation.
When Off is selected, operation will continue until it is
stopped with the Run/Stop key.
At the Temp setting, the unit will stop automatically
according to the increase in the vapor temperature.
See the description on the stop temperature (Stop Temp).
Setting of Auto Stop is made on the Setting screen.
Turn the dial key to move the cursor to [Auto Stop] and
press it to start setting.
(The indication color will change from light blue to yellow.)
Turn the dial key to change the value to the time you want to
set and press it to determine.
(Once a value is determined, its color will return from
yellow to light blue.)
* You can set Temp only when the Vapor temperature
sensor (optional) is used.
* When you have set Temp, the mark for
controlling with the vapor temperature sensor will be
displayed.
* You can quickly change the values by turning the dial
key faster.
(Except for during operation)
(4) Setting of a Stop Temperature (Stop Temp)
This setting will be valid when you have set the operation
stop condition (Auto Stop) to
Temp.
Setting range
:
0.1
~
15.0
℃
(unit:0.1)
Initial value
:
3.5
℃
You can stop operation of the unit automatically based on
the margin of temperature increase (Stop Temp) from the
vapor temperature if you use a vapor temperature sensor
(optional).
When a temperature measured on the vapor temperature
sensor reaches the sum of the vapor temperature and the stop
temperature (Stop Temp) operation will stop.
* Make this setting larger if operation stops too early
and make it smaller if operation stops too late.
-25-
■
Setting screen
■
Setting screen
■
Setting screen
℃
Time
Set Press
Stop Temp
Vapor
temperature
Pressure
Time
Leak
Menu
Run
Stop
Power
MANUAL
Standby
00:00:00
Esc
hPa
Leak
Menu
Run
Stop
Power
1013
Set Press
200
hPa
Δ P
Auto
hPa
Auto Stop Off
Leak
Menu
Run
Stop
Power
MANUAL
Standby
00:00:00
Esc
hPa
Leak
Menu
Run
Stop
Power
1013
Vapor 27.6
℃
Set Press
200
hPa
Δ P
Auto
hPa
Auto Stop
Off
Stop Temp
3.5
℃
Leak
Menu
Run
Stop
Power
MANUAL
Standby
00:00:00
Esc
hPa
Leak
Menu
Run
Stop
Power
1013
Vapor 27.6
℃
Set Press
200
hPa
Δ P
Auto
hPa
Auto Stop
Temp
Stop Temp
3.5
℃
℃
(2) Setting of hysteresis (control width) (ΔP
)
Setting range
:
Auto, 1
~
20hPa(unit:1)
Initial value
:
Auto
In the Auto mode, pressure is automatically controlled
with the 2% of the fixed-pressure (Set Press) set as ΔP.
Setting of ΔP is made on the Setting screen.
Turn the dial key to move the cursor to ΔP and press it to
start setting.
(The indication color will change from light blue to
yellow.)
Turn the dial key to change the value to the pressure you
want and press it to determine.
(Once a value is determined, its color will return from
yellow to light blue.)
* No setting is necessary for the inverter pump.
* When the vapor temperature
sensor is connected
* When the vapor temperature
sensor is connected