-42-
■
System Parameter 3 Setting screen
Leak
Menu
Run
Stop
Power
End
SYSTEM PARAMETER 3
Pressure Offset 0 980 hPa
Vapor Temp Offset 0.0 27.8
℃
Pressure Unit hPa
Temperature Unit
℃
User DATA BASE Reset
RESET
All DATA BASE Reset RESET
Memory Initialize INITIAL
Leak
Menu
Run
Stop
Power
(20) Initialize (Memory Initialize
)
This function resets all settings to the initial settings.
The user solvent database and the database will also be reset.
* Note any important parameter settings before initializing which
returns any values changed after purchase to the initial
settings.
Move the cursor to Memory Initialize and press the dial key to
display the [YES] and [NO] keys.
Press the [YES] key to initialize.
Press the [NO] key to cancel.
(18) Reset of the user solvent database (User DATA BASE Reset)
Resets the user solvent settings of the PROGRAM mode to the initial
settings.
Move the cursor to User DATA BASE Reset and press the dial key to
display the [YES] and the [NO] keys.
Press the [YES] key to reset.
Press the [YES] key to reset.
(19) Database reset
(
All DATA BASE Reset
)
This function resets the PROGRAM mode settings to the initial settings.
* Resetting the database also resets all settings changed in the
program mode including the user solvents to the initial settings. Be
sure to note any necessary parameter settings before reset.
Move the cursor to All DATA BASE Reset and press the dial key to
display the [YES] and the [NO] keys.
Press the [YES] key to reset.
Press the [YES] key to reset.
(16) Changing the pressure unit (Pressure Unit
)
Selectable unit
:
hPa, kPa, mbar, mmHg,Torr
Initial setting
:
hPa
You can set a pressure display unit.
(17) Changing the temperature unit (Temperature Unit)
Selectable unit:
℃
, K
Initial setting:
℃
You can set a temperature display unit.
System Parameter 3 Setting screen [SYSTEM PARAMETER 3]
* The System parameter 3 setting screen will not appear
during operation.
(14) Pressure Offset (Pressure Offset
)
Setting range
:
-30
~
30(unit:1)
Initial value
:
0 hPa
You can set display offset for a measured pressure.
(15) Vapor temperature offset (Vapor Temp Offset
)
Setting range
:
-3.0
~
3.0
℃
(unit:0.1)
Initial value
:
0
℃
You can set display offset for a measured vapor temperature.