-27-
■
Setting screen
■
Setting screen
Mark Press
Stop Temp
Set Temp
Leak
Menu
Run
Stop
Power
AUTO
Standby
00:00:00
Esc
Leak
Menu
Run
Stop
Power
1013
Δ P
Auto
hPa
Auto Stop Off
Mark Press
hPa
----
Leak
Menu
Run
Stop
Power
AUTO-2
Standby
00:00:00
Esc
Leak
Menu
Run
Stop
Power
1013
Mark Press
hPa
----
Vapor 27.6
℃
Δ P
Auto
hPa
Auto Stop
Off
Set Temp
Auto
℃
Adjust Temp
1.5
℃
Mark Count
1
Stop Temp
3.5
℃
(4) Setting of operation stop conditions (Auto Stop)
Setting range
:
Off, 1
~
999min(unit:1)
Initial value
:
Off
The unit will stop automatically when the set time has
elapsed since the start of operation.
When Off is selected, operation will continue until it is
stopped with the Run/Stop key.
*
You can quickly change the values by turning the
dial key faster.
(3) Setting of hysteresis (control width) (ΔP
)
Setting range
:
Auto, 1
~
20hPa(unit:1)
Initial value
:
Auto
In the Auto setting, automatic control is performed with 2%
of the target pressure set as ΔP.
Changing of ΔP is made on the Setting screen.
Turn the dial key to move the cursor to ΔP and press and
turn the dial key to select a value you want and then press it
to determine.
*
No setting is necessary for the inverter pump
.
You can switch to the AUTO-2 mode by using a vapor
temperature sensor (optional).
In addition to the functions of the AUTO mode, the following
functions are available:
・
Fixed-value control at the pressure (Mark Press)
determined automatically appropriately to the vapor
temperature (Set Temp).
・
Automatic stop based on the margin of temperature
increase (Stop Temp) from the vapor temperature (Set
Temp).
・
You can use mixed solvent by using an adjusted
temperature (Adjust Temp) and the count of adjustment
(Mark Count).
5-4-2
AUTO-2 mode
(1) [Hold] function
This is the same as for the AUTO mode.
(2) Adjusting of the automatically determined pressure (Mark
Press)
This is the same as for the AUTO mode.
(3) Setting of hysteresis (control width) (ΔP
)
This is the same as for the AUTO mode.
(4) Setting of operation stop conditions (Auto Stop)
Setting range
:
Temp, Off. 1
~
999min (unit:1)
Initial value
:
Off
Set conditions for automatic stop.
At the Temp setting, the unit will stop automatically
according to the increase in the vapor temperature.
See the description on the stop temperature (Stop Temp).
When Off is selected, operation will continue until it is
stopped with the Run/Stop key.
* You can quickly change the values by turning
the dial key faster.
Time
Vapor
temperature
Pressure
Time
Slope
start