-29-
The product has a database of 65 solvents in total including
55 kinds of solvents including pre-registered 10 solvents
(pre-registered solvents) and 10 user solvents which the
user can freely set.
You can freely change registered solvents with those in the
database.
After operation has started, slope control shall be carried
out between the slop start pressure (Slope Point) and
the fixed-pressure (Set Press) for the slope time (Slope
Time) and fixed-control will be performed at the fixed-
pressure.
You can stop the unit automatically based on the operation
time by setting an operation stop condition (Auto Stop).
You can stop the unit automatically based on the
temperature increase margin from the vapor temperature
by using a vapor temperature sensor (optional).
* The setting of the programs for 55 solvents has
been selected to realize the optimal condition at the
bath temperature of 40
℃
, the cooling water
temperature between 5 and 10
℃
, and the room
temperature of 25
±
3
℃
.
User solvents are allocated to No.56
~
65 in the database
and you can change their settings as well as set a solvent
name up to 18 characters. Characters available are
alphanumeric characters, space and symbols ( ,.-+*/() ).
Time
Slope Point
■
Control mode select screen
■
Solvent registration screen
Leak
Menu
Run
Stop
Power
PROGRAM
Standby
00:00:00
Esc
1013
hPa
Set Press
hPa
240
Leak
Menu
Run
Stop
Power
Acetone
Leak
Menu
Run
Stop
Power
DataBase
PROGRAM
Standby
00:00:00
End
hPa
Leak
Menu
Run
Stop
Power
1013
SELECT DATA 10
2 Aceton
7 Benzene
13 Chloroform
17 1,2-Dichloroethane
21 Dichloromethane
26 Diethyl ether
Acetone
No. of registered solvents
Set Press
Pressure
Slope Time
Auto Stop
5-5 Operations in the PROGRAM (Slope Control) mode
(1) Solvent select (Program Select)
Initial value: [Acetone]
You can select a solvent on the control mode select screen.
When you press the dial key, the cursor at the solvent name
will change to yellow. Turn the dial key to display the solvent
name you want and press it to determine.
The registered solvents at the factory shipping are as follows:
No
:
Solvent name
(
Slope Point→Set Press
:
Slope Time)
2
:
Acetone
(
350hPa→240hPa
:
3min
)
7
:
Benzene
(
160hPa→ 90hPa
:
2min
)
13
:
Chloroform
(
380hPa→200hPa
:
3min
)
17
:
1,2-Dichloroethane
(
180hPa→ 70hPa
:
3min
)
21
:
Dichloromethane
(
950hPa→480hPa
:
3min
)
26
:
Diethyl ether
(
800hPa→550hPa
:
3min
)
30
:
Ethyl acetate
(
266hPa→ 93hPa
:
2min
)
35
:
Hexane
(
280hPa→160hPa
:
2min
)
52
:
Toluene
(
90hPa→ 20hPa
:
4min
)
53
:
1,1,1-Trichloroethane
(
340hPa→160hPa
:
3min
)
You can select a solvent from the solvent registration screen as
well.
The selected solvent is indicated with a light blue cursor and
others are indicated with a black frame cursor.
Turn the dial key to move the cursor to the solvent you want
and then press it to determine.
* You can change each of the settings according to the
operating conditions.
See P.30 for how to change settings.
* See P.31 for how to change the registered solvents.