Causes and countermeasures of troubles
Symptom
Causes
Countermeasures
Power will not be turned on.
The LCD does not come on.
The power cable is not connected.
(Any of Diaphragm pump control
box PBX, inverter diaphragm
vacuum pump NVP-V, or DC
adaptor)
Connect the cable.
Power of the NVC-3000 is turned
OFF.
Press the power key to turn power ON.
Power is not supplied.
Turn the breaker on the distribution board
ON.
The fuse for PBX has blown.
Replace the fuse.
If the fuse is blown again soon after
replacement, stop operation immediately and
contact your dealer or our service center.
A malfunction occurred to the
control substrate.
Stop operation immediately and contact your
dealer or our service center.
A malfunction occurred to the LCD
indicator.
The vacuum pump will not
operate.
Power for PBX is not supplied.
Turn the breaker on the distribution board
ON.
Turn power for the vacuum unit ON.
The power of the vacuum pump is
OFF.
Turn the power switch of vacuum pump ON.
The power plug of the vacuum
pump has come off of the power
socket of the pump.
Turn the power switch OFF and insert the
power plug securely into the power socket of
the pump.
During leak and the control starting
pressure is not attained yet.
Continue operating as it is.
The connection cord is not inserted
into the outlet for PBX.
First turn power OFF and insert the
connection cord securely into the outlet.
A malfunction occurred to the
diaphragm vacuum pump.
Stop operation immediately and contact your
dealer or our service center.
There is no output at the outlet.
The inverter pump will not
operate.
Connection is not made properly
with the connection cord.
Connect the cable by referring to P.15
“When using the inverter diaphragm type
vacuum pump”.
Power of the inverter pump is
turned OFF.
Turn power of the inverter pump ON.
Any key operations are not
accepted.
Any dial operations are not
accepted.
A malfunction occurred to the
control substrate.
Stop operation immediately and contact your
dealer or our service center.
-46-