(8) Starting Step No
(
Start Step No
)
Setting range
:
1
~
(Setting for the finishing step)
Initial value
:
1
You set the start step No. for a program.
The range to execute is indicated with a red line
(
).
(9) Finishing Step No
(
Finish Step No
)
Setting range
:
(Setting for the start step)
~
99
Initial value
:
5
You can set the start step No. for a program
The range to execute is indicated with a red line
(
).
■
Program operation setting screen
-37-
(10)
Number of repetitions (Repeat)
Setting range
:
∞ , 1
~
999
Initial value
:
1
You can set a number of repetitions for a step program.
Your program will be repeated for the number of times you
have set.
When ∞ is set, the program will be repeated for the
unlimited times.
■
Program setting screen
■
Program operation setting screen
Leak
Menu
Run
Stop
Power
End
00:00:00
Display
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
P No
1
hPa
1013
Press(hPa)
Time (min)
ΔP (hPa)
Condition
Priority
Step No 1 2 3 4 5
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
00:00:00
End
Leak
Menu
Run
Stop
Power
1013
hPa
Start Step No
1
Finish Step No 5
Repeat 1
Auto Leak On
Auto Cleaning 3 min
Power Fail Cont
Back
P No
1
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
00:00:00
Esc
Leak
Menu
Run
Stop
Power
1013
hPa
Start Step No 1
Finish Step No 5
Repeat 1
Auto Leak
On
Auto Cleaning 3 min
Power Fail Cont
P No
1
After having completed setting of each step, make settings for
operation on the Program operation setting screen.
You can display the Program operation setting screen by
pressing the [Display] key on the Program setting screen.
You can make setting for the five Step programs.
separately.
(11) Auto Leak (Auto Leak) setting
Setting range:On, Off
Initial value
:
On
You can set On or Off for the release of decompression at
the end of operation.
When it is set to On, a series of operation "Pump output
Off→Auto leak" will be carried out automatically when
operation has completed.
【
Caution
】
Note that auto cleaning will not operate while Auto Leak
is set at Off.
In principle, set Auto Leak to On.
(12) Auto Cleaning (Auto Cleaning
)
setting
Setting range
:
Off, 1
~
10min(unit:1)
Initial value
:
3 min
You can set an Auto cleaning time for the pump and the
control solenoid valve after Auto Leak.
【
Caution
】
In principle, do not set Auto cleaning Off because it
purges liquefied steam from the control solenoid valve
or diaphragm vacuum pump. (Set a time.)