(8) Setting a count of adjustment (Mark Count)
Setting range: 1
~
10, Unlimited
Initial value
:
1
You can set a count of adjustment for mixed solvents.
Initial determination after slope control shall be the first count
for automatic determination of a pressure/.
At Unlimited, pressure will be adjusted based on the adjustment
temperature for an unlimited number of times.
* When pressure will not be adjusted for example, for a
single solvent, set this item to 1.
* When conditions for automatic stop are met before the
set count of adjustment is reached, operation will be
automatically stopped.
* When 2 or larger count is set for the count of
adjustment, the
mark will appear indicating that
control is being made with a sensor.
(5) Setting a vapor temperature (Set Temp)
Setting range
:
Auto, 0
~
150.0
℃
(unit:0.1)
Initial value
:
Auto
Pressure will be determined automatically according to the set
vapor temperature (Mark Press) and fixed-value operation will
be carried out.
In the Auto setting, pressure (Mark Press) will be determined
automatically as in the case of the AUTO mode.
* You cannot modify settings once slope operation has
started.
* When you have set a temperature, the mark
for controlling with the vapor temperature sensor will
be displayed.
* You can quickly change the values by turning the dial
key faster.
■
Setting screen
℃
-28-
■
Setting screen
■
Setting screen
*When the vapor temperature
sensor is connected
℃
Stop Temp
Set Temp
Adjust Temp
Mark Count=1
Mark Count=2
Leak
Menu
Run
Stop
Power
AUTO-2
Standby
00:00:00
Esc
Leak
Menu
Run
Stop
Power
1013
Mark Press
hPa
----
Vapor 27.6
℃
Δ P
Auto
hPa
Auto Stop
Temp
Set Temp
Auto
℃
Adjust Temp
1.5
℃
Mark Count
1
Stop Temp
3.5
℃
℃
Leak
Menu
Run
Stop
Power
AUTO-2
Standby
00:00:00
Esc
Leak
Menu
Run
Stop
Power
1013
Mark Press
hPa
----
Vapor 27.6
℃
Δ P
Auto
hPa
Auto Stop
Off
Set Temp
Auto
℃
Adjust Temp
1.5
℃
Mark Count
1
Stop Temp
3.5
℃
Leak
Menu
Run
Stop
Power
AUTO-2
Standby
00:00:00
Esc
Leak
Menu
Run
Stop
Power
1013
Mark Press
hPa
----
Vapor 27.6
℃
Δ P
Auto
hPa
Auto Stop
Off
Set Temp
25.0
℃
Adjust Temp
1.5
℃
Mark Count
2
Stop Temp
3.5
℃
℃
℃
(6) Setting of a Stop Temperature (Stop Temp)
This setting will be valid when you have set the operation stop
condition (Auto Stop) to Temp.
Setting range
:
(Adjust Temp+0.1)
~
15.0
℃
(unit:0.1)
Initial value
:
3.5
℃
When a temperature measured on the vapor temperature sensor
reaches the sum of the vapor temperature (Set Temp) and the
stop temperature (Stop Temp) operation will stop.
* Make this setting larger if operation stops too early and
make it smaller if operation stops too late.
(7) Setting of a adjusting temperature (Adjust Temp)
Setting range
:
0.1
~
(Stop Temp-0.1)
℃
(unit:0.1)
Initial value
:
1.5
℃
When one of mixed solvents has been used up, you can
resume decompression automatically and determine a
pressure for the next solvent automatically.
When a measured vapor temperature reaches the sum of
the vapor temperature
(
Set Temp) and the adjustment
temperature (Adjust Temp) decompression resumes.
Time
Mark Press
(1set session)
Vapor
Temperature
Pressure
Time
Mark Press
(2nd session)