There are five programs up to 99 steps each
to be used as a programed vacuum controller which allows
programming of fixed-value control, slope control and
release to the atmosphere.
Fixed-value control will be made for the set time .
In slope control, decompression will be made at the set time
from the slope start
pressure
(
set pressure of the
previous step
)
to the set pressure.
You can set a step program on the Program setting screen.
The items and the initial values of the steps are as follows:
Set pressure
(Press)
:
VAC [hPa]
Step operation time
(Time)
:
0 [min]
Hysteresis
(ΔP)
:
Auto
Control condition select
(Condition)
:
Step
Selecting a priority condition
(Priority)
:
Press
* Setting the control condition select (Condition) to
Step starts the fixed value control and setting it
Slope start slope control.
* Setting priority condition select (Priority) to Press
continues the present step until the set pressure is
attained after the set time has elapsed.
When it is set to Time, the process will move to the
next step when the step operation time has
elapsed.
* For the inverter pump, you do not need to set a
hysteresis (ΔP).
You can set program operations on the Program setting
screen. The items and the initial values are as follows:
Starting Step No.
(Start Step No)
:
1
Finishing Step No.
(Finish Step No)
:
5
Number of repetitions
(Repeat)
:
1
Auto Leak
(Leak)
:
ON
Auto Cleaning
(Auto Cleaning)
:
3 [min]
Recovery from Power Failure (Power Fail)
:
Cont
-34-
StepNo2
ΔP
Time
Pressure
StepNo1
StepNo3
StepNo4
StepNo5
StepNo2 StepNo3 StepNo4 StepNo5
5-6 STEP PROGRAM(Step Program)
Mode operations