(6) Copy
[Copy]
You can copy the settings for the previous step when make
settings for the Step No.2 and following steps.
Pressing the [Copy] key after staring setting, the settings for
the previous step will be copied for the item.
* At this stage, the settings are only copied and not
determined.
Press the dial key to determine.
-36-
(5) Selecting a priority condition (Priority)
Setting conditions
:
Press, Time
Initial value
:
Press
Press
The present step will continue until the set pressure
is attained after the set time has elapsed by putting
priority on the pressure.
Time
The step will move to the next step after the set time
has elapsed with the time prioritized.
■
Program setting screen
Leak
Menu
Run
Stop
Power
00:00:00
Esc
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
P No
1
hPa
1013
Press(hPa)
Time (min)
ΔP (hPa)
Condition
Priority
Step No 1 2 3 4 5
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
Leak
Menu
Run
Stop
Power
00:00:00
Esc
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
P No
1
hPa
1013
Press(hPa)
Time (min)
ΔP (hPa)
Condition
Priority
Step No 1 2 3 4 5
520
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
Copy
Leak
Menu
Run
Stop
Power
00:00:00
Display
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
P No
1
hPa
1013
Press(hPa)
Time (min)
ΔP (hPa)
Condition
Priority
Step No 1 2 3 4 5
520
0
Auto
Step
Press
520
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
End
Leak
Menu
Run
Stop
Power
00:00:00
Esc
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
P No
1
hPa
980
Press(hPa)
Time (min)
ΔP (hPa)
Condition
Priority
Step No 1 2 3 4 5
520
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
Copy
Leak
Menu
Run
Stop
Power
00:00:00
Ins
Leak
Menu
Run
Stop
Power
STEP PROGRAM
Standby
P No
1
hPa
1013
520
0
Auto
Step
Press
520
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
VAC
0
Auto
Step
Press
Del
Press(hPa)
Time (min)
ΔP (hPa)
Condition
Priority
Step No 1
2
3 4 5
(7) Insert [Ins], deletion [Del] and movement of display of a step
You can insert or delete steps.
Turn the dial key to move the cursor to a step No.
Pressing the dial key changes the color of the step No. and its
settings to yellow.
Pressing the [Ins] key inserts a new step before the current
step.
(Initial value: VAC, 0, Auto, Step, Press)
【
Caution
】
You cannot insert a step when a setting is already
entered in Step No.99.
Pressing the [Del] key deletes the current step.
Turning the dial key moves the step at the cursor position.
Pressing the dial key returns to the original indication.