Product name
Vacuum controller
Model
NVC-3000
Operating ambient temperature range
5
~
35
℃
Operating ambient humidity range
30
~
80%RH(no condensation)
External dimensions (W x D x H) mm
* External dimensions in parentheses ( )
include protrusions.
138.7(150)
×
50(114.6)
×
114.2
Weight: kg
Approx. 670g
Power input, rated power supply
DC24V
375mA
-3-
※
1 Setting of the hysteresis (ΔP) is unnecessary when you use an inverter pump.
※
2 Percentage of a target gradient pressure.
※
3 CV-11 (For a condensation unit of 1
~
5 liters)
CV-12
(
For a condensation unit of 5
~
10 liters
)
・
When you use an inverter diaphragm type vacuum pump,
connect it with the model NVC-3000 with a communication cable.
Vacuum is controlled by continuously varying the number of
rotations of the pump motor by using control signals for the model
NVC-3000.
Vacuum level is controlled by increase the number of rotations
immediately after starting vacuum control and decreasing the
number of rotations at around the set pressure (stops below the
lowest number of pump motor rotations).
・
When you use a diaphragm type vacuum pump (constant
rotation speed)
, vacuum is controlled by connecting the model
NVC-3000, the diaphragm control box PBX, and the NVC control
solenoid valves CV-11
・
12 models.
Model PBX receives control signals from NVC-3000 to control the
vacuum level.
・
Open/close the solenoid valve (vacuum line)
・
Start/stop the pump motor
※
1
You can set a hysteresis ΔP(Pressure control width by
opening/closing the solenoid valve).
2-3 Vacuum control (differences among pumps)
Time
ΔP
Pressure
Set pressure
Pressure
Time
Vacuum control start
Vacuum control start
Set pressure
Product name (Sold separately)
Diaphragm pump control box
Type
PBX
Standards
Pump output (outlet) MAX6A
Operating environmental temperature range
5
~
35
℃
Operating environmental humidity range
30
~
80%RH(no condensation)
External dimensions (W
×
D
×
H)
mm
63
×
63
×
148
Weight kg
Approx. 570g
Power input, rated power supply
100V
~
240V 1A
(
NVC-3000 is connected
)