The software recognises missing slide IDs. It is not possible to start the
worklist. Each Slide ID line needs to contain an entry.
Figure 7-44: Missing Slide ID
The Slide ID of every slide is visible on the top of the print out.
Figure 7-45: Print-out with Slide ID
7.4.3 Loading the Sprinter XL for ELISA
Proceed as follows:
Was daily maintenance performed? If not, follow the steps described in
chapter 7.2.
Check the containers for system liquid and liquid waste. Fill them as necessary
or empty them and ensure the tubes fit tightly.
Place the appropriate wash buffer tubes into the filled wash buffer containers.
Start the worklist created according to chapter 7.3 by clicking on
the
Run
button.
Empty Slide IDs
Print out