- 48 -
7 Operating instructions
M e d i z i n i s c h e
L a b o r d i a g n o s t i k a
A G
EUROIMMUN
The same Slide ID cannot be assigned twice. If during manual reading process
the Slide ID is assigned twice, the following error message appears:
Figure 7-42: Twice assigned Slide ID during manual reading in via handscanner
The message contains the ID assigned in duplicate and the
associated position
on the slide rack.
After the message is confirmed, the ID assigned in duplicate is automatically
replaced by a new predefined ID. Afterwards reading process can be repeated.
In case the Slide ID is assigned twice on different slide racks, the following
error message is displayed after clicking the button
“start”.
Figure 7-43: Twice assigned Slide ID on different rack
The message contains the ID assigned in duplicate and the associated slide
rack. The worklist cannot be started. The twice assigned Slide ID needs to be
corrected.
Twice assigned
Slide ID