- 68 -
10 Error remedies
M e d i z i n i s c h e
L a b o r d i a g n o s t i k a
A G
EUROIMMUN
10 Error
remedies
10.1
Error message: No Liquid found
If the filling level of a sample or a reagent is not detected when the liquid is
drawn up by the needle, the following error message appears with different
fault remedy possibilities:
Figure 10-1:
No liquid found
Click on
Retry
to repeat the process.
Select the
Ignore
option to switch off the liquid level detection
sensor. The needle moves until just over the tube bottom and
starts aspirating without checking whether it is in liquid.
Click on
Abort
to cancel the worklist.
10.2
Error message: Movement of (33) blocked
If a pipetting needle meets resistance during the test run, the following
message appears:
Figure 10-2:
Movement blocked
It is not possible to continue the test run. The test run must be cancelled and
restarted.