U2829 Series Operation Manual Chapter 7 Command Reference
93
Only when the instrument compare function is set as ON, <bin No.> data can be
displayed.
The output format of <bin No.> data applies 2 to 3 digits ASCII: SN or SNN (S:
+/_, N: from 0 to 9).
On list sweep display page, the ASCII data output format is shown as
below, that is, the return-circuit replaces sweep point number.
,
SN.NNNNNESNN
,
SN.NNNNNESNN
,
SN
,
SN NL^END
<DATA A> <DATA B> <Status> <Judge>
Figure 6 ASCII format 2 (list sweep)
Where,
Descriptions for <DATA A>, <DATA B>, <Status> are the same described before.
<Judge> format is as below:
<Input/Output> format: The data displays the compare result of the list sweep..
Data Result
-1
0
+1
low
pass
high
When the compare function of the list sweep measurement is turned off, the
output result of <Input/Output> is 0.
<Input/Output> data output format applies 2-digits ASCII format: SN (S: +/_, N:
from 0 to 1)
On transformer single-machine measurement display page, ASCII data output
format is as below:
SN
,
SN.NNNNNESNN
,
SN.NNNNNESNN
,
SN NL^END
<Serial number><DATA A> <DATA B> <Status>
<SN> is described as below:
SN Description
1 TURN
2 Zx
3 Acr
4 Lx
5 LK
6 DCR
Description of <DATA A> is the same as that described above. Secondary
parameter of <DATA B> has two formats:
When the measurement parameter is the turn ratio (TURN), secondary
parameter <DATA B> is the phase and input format is + (in-phase), -
(anti-phase);