U2829 Series Operation Manual Chapter 6 Performance and Test
72
capacitor
1000pF
D is known
10000pF
10nF
0.1uF
1uF
2
AC
standard resistor
10
Ω
0.02
%
100
Ω
1k
Ω
10k
Ω
100k
Ω
3
DC
standard resistor
0.1
Ω
0.02
%
1
Ω
10
Ω
100
Ω
1k
Ω
10k
Ω
100k
Ω
4 Standard
inductor
100
μ
H
0.02%
1mH
10mH
100mH
5 Frequency
counter
(
0
~
1000
)
MHz
6 Digital
Multimeter
0.5
%
7
Insulation resistance meter
500V 10 levels
8 Hipot
Tester
0.25kW
(
0
~
500
)
V
6.6.3
Function check
Ensure function keys, displayer and terminal etc. can work normally.
6.6.4
Test signal level
Adjust multimeter in AC voltage range, where one test cable is connected to H
CUR
and the other is connected to ground terminal. Change level as: 10mV, 20mV,
100mV, 200mV, 1V, 2V, the reading should meet the demand of test signal level in
this chapter.
6.6.5
Frequency
Connect frequency meter to ground terminal. The test terminal of the frequency
meter is connected with H
CUR.
Change the frequency as: 20Hz, 100Hz, 1kHz,