U2829 Series Operation Manual Chapter 7 Command Reference
90
FAST
APERture MEDium [,<value>]
SLOW
Where,
FAST: 30 times/sec
MEDium: 10 times/sec
SLOW: 2 times/sec
<value> 1 to 255 in NR1
For example: WrtCmd (“APER MED, 55”)
Query syntax: APERture?
Return format:
FAST
MED , <NR1><NL^END>
SLOW
7.1.12
TRIGger subsystem commands
The TRIGger subsystem commands are mainly used to set the instrument trigger
source, trigger delay and trigger measurement.
Command tree:
TRIGger
[:IMMediate]
:SOURce INTernal
EXTernal
BUS
HOLD
:DELay <value>
MIN
MAX
The TRIGger[:IMMediate] command is used to trigger a test.
Command syntax: TRIGger[:IMMediate]
For example: WrtCmd(“TRIG”)
The TRIGger:SOURce command is used to set the trigger source mode. The
TRIGger:SOURce? query returns the current trigger source mode.
Command syntax:
INTernal
TRIGger:SOURce EXTternal
BUS
HOLD
Where,
INTernal The default trigger mode.