U2829 Series Operation Manual Chapter 3 Introduction to [DISP]
14
U2829 has 8 AC measurement ranges: 10
Ω
, 100
Ω
, 300
Ω
, 1k
Ω
, 3k
Ω
, 10k
Ω
, 30k
Ω
,
100k
Ω
.
U2829 has 10 DCR measurement ranges: 100m
Ω
, 1
Ω
, 10
Ω
, 100
Ω
, 300
Ω
, 1k
Ω
,
3k
Ω
, 10k
Ω
, 30k
Ω
, 100k
Ω
.
Operation steps for setting test range:
1)
Move the cursor to the range zone, the following soft keys will be displayed:
AUTO The soft key is used to set the range mode to AUTO.
HOLD The soft key is used to switch the AUTO mode to the HOLD mode.
When the range mode is set to HOLD, the range will be locked in the
current measurement range. The current measurement range will be
displayed in the range zone.
↑
(+) The soft key is used to increase the range under HOLD mode.
↓
(-) The soft key is used to decrease the range under HOLD mode.
2)
Use soft keys to set measurement range.
3.1.3
Test frequency
The measurement range of U2829 ranges from 20Hz to 500KHz with an increase
or decrease of 0.01Hz. When the test function is set as DCR, the
FREQ
zone will
display “---”.
Operation steps for setting test frequency:
U2829 provides two methods to set measurement frequency. The first one is to
use soft keys and the other one is to input data by using numeric keys.
1)
Move the cursor to the
FREQ
zone, the following soft keys will be displayed.
↑
(++)
This is a coarse adjustment soft key used to increase the frequency. Press
this key, the frequency will change as the following sequence: 20Hz,
100Hz, 1kHz, 10kHz, 100kHz, 200kHz.
↑
(+)
This is a fine adjustment soft key used to increase the frequency. Press
this key, the frequency will switch between the following ones:
20 Hz 100 Hz 1 kHz 10 kHz 100 kHz
25 Hz 120 Hz 1.2 kHz 12 kHz 120 kHz
30 Hz 150 Hz 1.5 kHz 15 kHz 150 kHz
40 Hz 200 Hz 2 kHz 20 kHz 200 kHz
50 Hz 250 Hz 2.5 kHz 25 kHz 250 kHz
60 Hz 300 Hz 3 kHz 30 kHz 300 kHz
80 Hz 400 Hz 4 kHz 40 kHz 400 kHz
500 Hz 5 kHz 50 kHz 500 kHz
600 Hz 6 kHz 60 kHz
800 Hz 8 kHz 80 kHz
↓
(-)
This is a fine adjustment soft key used to decrease the frequency. The