U2829 Series Operation Manual Chapter 5 Execute LCR Operation and Some Examples
57
U2829501
7)
If the test result is obviously incorrect, please check the following items.
a)
Check the tested inductance is in good connection with the test fixture or not.
b)
Check the test fixture is in good connection with the test terminals of the
instrument or not.
c)
Redo the open/short correction.
NOTE:
When the sweep open/short correction is used, the point-frequency
correction function should be set as OFF. Refer to Correction operation in this
chapter.
5.5
Operation example of testing capacitance by
multi-frequency list sweep
Test condition
Function: Cp-D
Level: 1Vrms
Other parameters
Frequenc
y
Compare
parameter
Low limit
High limit
1kHz Cp(capacitance)
325.0nF
333.0nF
10kHz
D
(
Dissipation
)
0.0001 0.0003
100kHz
D
(
Dissipation
)
0.0060 0.0100
Beep: LONG HIGH
Alarm mode: OUT
Operation steps
1)
Turn on the instrument.
2)
Set basic parameters.