U2829 Series Operation Manual Chapter 7 Command Reference
92
WrtCmd (“FETC?”)
U2829 applies ASCII to delivery result, details are follows.
On measurement display page, bin NO. display page, bin count display page,
ASCII data output format are described as below:
SN.NNNNNESNN
,
SN.NNNNNESNN
,
SN
,
SN or SNN NL^END
<DATA A>
<DATA B>
<Status>
<BIN number>
Where,
<DATA A>, <DATA B> format: <DATA A> (primary measurement data), <DATA
B> (secondary measurement data)
12-digits ASCII format are as below:
SN.NNNNNESNN
(S:+/-,N: from 0 to 9, E: Exponent Sign)
Status Description
-1
0
+1
+2
+3
+4
(In data buffer memory) no data
Common measurement data
Analog LCR unbalance
A/D converter is not working.
Signal source is over loading.
Constant voltage cannot be adjusted.
<status> format: When above measurement data is used, <status> data will
display measurement status.
The output format of the <Status> display data uses 2-digits ASCII: SN (S: +/-,
N: from 0 to 4)
NOTE: When <status> is -1, +1 or +2, the measurement data is 9.9E37.
When <status> is 0, +3 or +4, the real measurement data is beyond the
limits.
Data Sort result
0
+1
+2
+3
+4
+5
+6
+7
+8
+9
+10
Out of tolerance
Bin 1
Bin 2
Bin 3
Bin 4
Bin 5
Bin 6
Bin 7
Bin 8
Bin 9
Auxiliary bin
<Bin No.> format: The data displays the sorting results of the displayed bin,
shown as above.