U2829 Series Operation Manual Chapter 7 Command Reference
99
Command syntax: CORRection:SPOT<n>:OPEN
Where,
<n>:
1 refers to frequency spot 1.
2 refers to frequency spot 2.
3 refers to frequency spot 3.
For example: WrtCmd (“CORR:SPOT1:OPEN”) Execute open correction for
frequency spot 1.
The CORRection:SPOT<n>:SHORt command is used to execute short correction
for specific frequency spots (frequency 1, frequency 2 and frequency 3).
Command syntax: CORRection:SPOT<n>:SHORt
Where,
<n>:
1 refers to frequency spot 1.
2 refers to frequency spot 2.
3 refers to frequency spot 3.
For example: WrtCmd (“CORR:SPOT1:SHOR”) Execute short correction for
frequency spot 1.
The CORRection:SPOT<n>:LOAD:STANdard command is used to set the
standard reference of specific frequency spots (frequency spot 1, frequency spot2
and frequency spot 3 which are taking load correction). The
CORRection:SPOT<n>:LOAD:STANdard query returns the current standard
reference of specific frequency spot.
Command syntax: CORRection:SPOT<n>:LOAD:STANdard <REF. A><REF. B>
Where,
<n>:
1 refers to frequency spot 1.
2 refers to frequency spot 2.
3 refers to frequency spot 3.
<REF. A> can be NR1, NR2 or NR3 data format and taken as the standard
reference of the primary parameter.
<REF. B> can be NR1, NR2 or NR3 data format and taken as the standard
reference of the secondary parameter.
For example: WrtCmd (“CORR:SPOT1:LOAD:STAN 100.7, 0.0002”)
Query syntax: CORRection:SPOT<n>:LOAD:STANdard?
Return format: <NR3><NL^END>
The CORRection:USE command is used to set the channel number of
multi-channel correction. The CORRdction:USE? query returns the current used
channel number.
Command syntax: CORRection:USE <channel number>