U2829 Series Operation Manual Chapter 3 Introduction to [DISP]
25
———————————————————————————————————————
Note: Bias current isolation function (ISO) can only be set as ON after installing
bias board. After bias current isolation function is turned on, the test accuracy will
be influenced, so under the condition of low frequency and bias current, the bias
current isolation function should be set as OFF.
———————————————————————————————————————
Operation steps for setting the bias current isolation function
(Bias board
has been installed)
Execute the following steps and set the bias current isolation function as ON or
OFF.
1)
Move the cursor to
DCI ISO
zone, the following soft keys will be displayed.
ON
OFF
2)
Press
ON
to turn on the bias current isolation function.
3)
Press
OFF
to turn off the bias current isolation function.
3.4.4
Average
The AVERAGE function can calculate the average value of two or more test results.
The average times can be set from 1 to 255 with an increase or decrease of 1.
Operation steps for setting test average times.
1)
Move the cursor to the
AVG
zone, the following soft keys are displayed.
↑
(
+
)
This key is used to increase the average times.
↓
(
-
)
This key is used to decrease the average times.
2)
Use above soft keys to set the average times or use numeric keys and
[
ENTER
] input average times directly.
3.4.5
Level monitor function
The level monitor function can monitor the real voltage across DUT or real current
through DUT. The monitored voltage value is displayed in
VAC MON
zone on
<
MEAS DISPLAY
> page while the monitored current value is in
IAC MON
zone.
———————————————————————————————————————
Note: The correction function can influence the level monitor function, so when
the correction data changes the level monitor value will change. When the
correction is switched between OPEN or SHOR or LOAD, the level monitor value
will be influenced as well
———————————————————————————————————————
Operation steps for setting the level monitor function
Execute the following operation steps to set the level monitor function as ON or