U2829 Series Operation Manual Chapter 3 Introduction to [DISP]
21
Fail mode(
FAIL MODE
)
There are 2 zones on this page:
LIST SWEEP DISP
and
mode
. List sweep points
cannot be set on this page but can be set on <
LIST SWEEP SETUP
>.
3.3.1
Sweep mode
The list sweep function of U2829 can make automatic sweep test for up to 10
points’ test frequencies, test levels or DC bias. Two sweep modes are available on
U2829: SEQ and STEP. In SEQ mode, each press of [
TRIG
] will direct U2829 to
automatically test all list sweep test points. In STEP mode, each press of [
TRIG
]
will direct U2829 to test one list sweep point.
———————————————————————————————————————
NOTE: When the trigger mode is set to INT, sweep test modes of SEQ and STEP
will not be controlled by [
TRIG
].
When the trigger mode is set to MAN, [
TRIG
] can be used to trigger the list sweep
test.
———————————————————————————————————————
Operation steps for setting the list sweep mode:
Set the sweep mode on the <
LIST SWEEP
> page as
SEQ
or
STEP
.
1)
On the <
LIST SWEEP
> page, move the cursor to the
SWEEP MODE
zone,
the following soft keys will be displayed:
SEQ
STEP
2)
Press
SEQ
to set the sweep mode as sequential sweep test mode.
3)
Press
STEP
to set the sweep mode as single step sweep test mode.
Operation steps for setting the list sweep fail mode:
Set the sweep fail mode on the <
LIST SWEEP
> page as
STOP
or
CONTINUE
.
4)
On the <
LIST SWEEP
> page, move the cursor to the
FAIL MODE
zone, the
following soft keys will be displayed:
STOP
CONTINUE
PAUSE
RETRY
3.3.2
FREQ (Hz)
This zone shows the currently swept parameter mode and its unit. What are right
below this item are parameters of the sweep list.
3.3.3
Ls[H] Q[ ]
This zone is the currently swept Function parameter and its unit. What are right