U2829 Series Operation Manual Chapter 5 Execute LCR Operation and Some Examples
58
a)
Press [MEAS] to enter into the <MEAS DISPLAY> page.
b)
The
FUNC A-B
zone is currently displayed as Cp-D and the Level zone is
1.000V.
c)
Press [SETUP] to enter into the <MEAS SETUP> page, meanwhile the
following soft keys will be displayed in the soft key zone: MEAS SETUP,
CORRECTION, LIMIT TABLE, LIST SETUP and GRAPH SETUP.
d)
Press LIST SETUP to enter into the <LIST SETUP> page.
e)
Move the cursor to the
FUN
zone of sweep point 0, then the soft key zone
will be displayed as OFF,Ls,Lp,Cs,Cp,MORE. Press Cp to set the parameter
zone of sweep point 1.
f)
Move the cursor to
FREQ
zone of sweep point 1, then this zone will
displayed as 1.00000k.
g)
Move the cursor to
LEVEL
zone of sweep point 1, then this zone will
displayed as 1.0000.
h)
Move the cursor to
LMT:LOW
zone of sweep point 1, then this zone will
displayed as 0.00000pF.
i)
Press [3][2][5], +325 will be displayed in the prompt information zone
and the following available units will be displayed in the soft key zone: p,
n,
μ
, m. Press n, this zone will be changed as 325.000n.
j)
Move the cursor to
LMT:HIGH
zone of sweep point 1, then this zone will
displayed as 0.00000pF.
k)
Press [3] [3] [3], +333 will be displayed in the prompt information zone
and the following units will be displayed in the soft key zone: p, n,
μ
, m.
Press n, this zone will be changed as 333.000n.
l)
Move the cursor to the
FUN
zone of sweep point 2, then the soft key zone
will be displayed as OFF,Ls,Lp,Cs,Cp,MORE.
m)
Press MORE, then the soft key zone will display LESS,Rs,Rp,R,ESR,MORE.
n)
Press MORE, then the soft key zone will display LESS,D,Q, MORE, Press D
to set the parameter zone of sweep point 2.
o)
Move the cursor to
FREQ
zone of sweep point 2, then this zone will
displayed as 1.00000k.
p)
Press [10], +10 will be displayed in the prompt information zone and the
soft key zone will display the following available units: Hz, kHz and MHz.
Press kHz, this zone will be displayed as 10.0000k.
q)
Move the cursor to
LEVEL
zone of sweep point 2, then this zone will
displayed as 1.0000.
r)
Move the cursor to
LMT:LOW
zone of sweep point 2, then this zone will
displayed as 0.00000p.
s)
Press [0][.][0][0][0][8], +0.0008 will be displayed in the prompt
information zone and the soft key zone will display the following available
units: p, n,
μ
, m. Press [ENTER], this zone will be changed as 0.00080.
t)
Move the cursor to
LMT:HIGH
zone of sweep point 2, then this zone will
displayed as 0.00000p.
u)
Press [0][.][0][0][2], +0.002 will be displayed in the prompt information