U2829 Series Operation Manual Chapter 3 Introduction to [DISP]
39
the data, you can use (
p
,
n
,
µ
,
m
,
k
,
M
,
*1
) to replace
[ENTER]
to input the
limit value. When [ENTER] is used to input the limit value, the default unit is
the same as that being input last time. When pressing
*1
, the default unit will
be F, H or
Ω
.
8)
After inputting the low limit of BIN 1, the cursor will automatically move to
HIGH
of BIN 1. Input the high limit of BIN 1.
9)
The cursor will automatically move
HIGH
of BIN 2. For the limit mode is
sequential mode, the low limit of BIN 2 will be the high limit of BIN 1. Input
the high limit of BIN 2.
10)
Repeat step 9 until the high limit of BIN 9 is input. Then the cursor will
automatically move to
LOW
of 2
nd
. Input the low limit of the secondary
parameter.
11)
The cursor will automatically move to
HIGH
of 2
nd
. Input the high limit of the
secondary parameter.
3.7
<LIST SETUP>
Press
[SETUP]
and then
LIST SETUP
to enter into the
<LIST SETUP>
page as
shown below.
U2829309
The list sweep function of U2829 can perform auto sweep test for the test
frequency, test level or bias voltage of 10 points. On
<LIST SETUP>
page, the
following list sweep parameters can be set.
Sweep mode (
SWEEP MODE
)
Sweep parameter setup (frequency
[Hz]
, level
[V]
, level
[I]
, bias
[V]
, bias
[I]
)