Guidance and manufacturer's declaration - electromagnetic immunity
The O-scan system is intended for use in the electromagnetic environment specified below. The customer or
the user of the system should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
level
Compliance
Level
Electromagnetic environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
r
8 kV contact
r
15 kV air
r
8 kV contact
r
15 kV air
Floors should be wood, concrete, ceramic tile or
covered with synthetic material. In any case the
minimum relative humidity shall be at least 45%.
The PC unit and the monitor are in compliance
with the immunity levels of their applicable
international EMC standards.
It is however suggested to observe the following
caution: when the O-scan system is in operation,
do not touch metallic parts of the PC or monitor,
and in particular connectors located on the
relative rear panels.
Electrical fast
transient/burst
IEC 61000-4-4
r
2 kV for power
supply lines
r
1 kV for
input/output lines
r
2 kV for power
supply lines
r
1 kV for
input/output lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
r
2 kV differential
mode
r
2 kV common
mode
r
1 kV differential
mode
r
2 kV common
mode
Mains power quality should be that of a typical
commercial or hospital environment.
Содержание O-scan eXP
Страница 1: ......
Страница 2: ......
Страница 3: ...O scan User Manual 350004900 Rev 07 Esaote S p A Genoa...
Страница 9: ...350004900 Rev 07 7 10...
Страница 10: ...8 10...
Страница 12: ...10 10...
Страница 20: ...8 8...
Страница 32: ...8 8 Chapter 2...
Страница 51: ...350004900 Rev 07 19 48 Instructions for Safe Use fig 3 1 Top view...
Страница 52: ...20 48 Chapter 3 fig 3 2 Side view...
Страница 80: ...48 48 Chapter 3...
Страница 102: ...22 22 Chapter 4...
Страница 105: ...350004900 Rev 07 3 20 Technical Description fig 5 1 Equipment components...
Страница 122: ...20 20 Chapter 5...
Страница 126: ...4 20 Chapter 6 fig 6 3 The O scan Premium and O scan eXP patient seat with armrests fig 6 4 Armrest detail...
Страница 142: ...20 20 Chapter 6...
Страница 155: ...350004900 Rev 07 7 8 Coils fig 8 5 Spherical uniform phantom fig 8 6 Cylindrical uniform phantoms...
Страница 156: ...8 8 Chapter 8 fig 8 7 Geometric phantom fig 8 8 Centering support for the spherical phantom diameter 140 mm...
Страница 160: ...4 10 Chapter 9 fig 9 4 Lumbar support cushion fig 9 5 Positioning of the lumbar support cushion on the seat...
Страница 161: ...350004900 Rev 07 5 10 Soft cushion Set fig 9 6 Use of lumbar support cushion...
Страница 184: ...10 10 Chapter 11...
Страница 232: ...4 4 Chapter 17...
Страница 236: ...4 4 Chapter 18...