![Esaote O-scan eXP Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/esaote/o-scan-exp/o-scan-exp_user-manual_2425230055.webp)
350004900 Rev. 07
23
/ 48
•
•
•
•
•
•
Instructions for Safe Use
.
fig. 3.4 - Regions of maximum magnetic field and magnetic field gradient
The values measured in the defined regions are as follows:
☛
Maximum main magnetic field: B0 = 1.1027 mT
☛
Maximum spatial gradient of main magnetic field: 9,2398 T/m
☛
Maximum result: (B0) x (spatial gradient B0) = 10.1887 mT
2
/m
Personnel and patient exposure to the time-varying magnetic field
A time-varying magnetic field induces an electric field
E
according to
Faraday's Law. The gradient coils in MR systems generate a time-varying
magnetic field, and therefore the body of the patient is exposed to an
induced electric field. This electric field can stimulate the heart, the
peripheral nervous system and the muscles, with thresholds that vary
according to frequency. The stimulation of the peripheral nervous system
(PNS) is induced by the minor gradient output values with respect to the
values required for cardiac stimulation. Therefore protection of the patient
from stimulation of the peripheral nervous system also serves to prevent
cardiac stimulation (CS).
Содержание O-scan eXP
Страница 1: ......
Страница 2: ......
Страница 3: ...O scan User Manual 350004900 Rev 07 Esaote S p A Genoa...
Страница 9: ...350004900 Rev 07 7 10...
Страница 10: ...8 10...
Страница 12: ...10 10...
Страница 20: ...8 8...
Страница 32: ...8 8 Chapter 2...
Страница 51: ...350004900 Rev 07 19 48 Instructions for Safe Use fig 3 1 Top view...
Страница 52: ...20 48 Chapter 3 fig 3 2 Side view...
Страница 80: ...48 48 Chapter 3...
Страница 102: ...22 22 Chapter 4...
Страница 105: ...350004900 Rev 07 3 20 Technical Description fig 5 1 Equipment components...
Страница 122: ...20 20 Chapter 5...
Страница 126: ...4 20 Chapter 6 fig 6 3 The O scan Premium and O scan eXP patient seat with armrests fig 6 4 Armrest detail...
Страница 142: ...20 20 Chapter 6...
Страница 155: ...350004900 Rev 07 7 8 Coils fig 8 5 Spherical uniform phantom fig 8 6 Cylindrical uniform phantoms...
Страница 156: ...8 8 Chapter 8 fig 8 7 Geometric phantom fig 8 8 Centering support for the spherical phantom diameter 140 mm...
Страница 160: ...4 10 Chapter 9 fig 9 4 Lumbar support cushion fig 9 5 Positioning of the lumbar support cushion on the seat...
Страница 161: ...350004900 Rev 07 5 10 Soft cushion Set fig 9 6 Use of lumbar support cushion...
Страница 184: ...10 10 Chapter 11...
Страница 232: ...4 4 Chapter 17...
Страница 236: ...4 4 Chapter 18...