350004900 Rev. 07
3
/ 10
•
•
•
•
•
•
Hand-W
rist examinations
Hand Examinations: Patient Positioning Procedure
DPA Hand/Wrist Coil 2 is recommended for examinations of the hand on
standard or small size patients.
DPA Hand/Wrist Coil 3 is recommended for examinations of the hand on
large size patients.
1
Check that there are no obstacles in the patient seat movement area.
2
Ensure that there is no coil in the gantry and that the cavity is totally
free.
3
After releasing the brakes, place the patient's seat in the position
parallel to the magnetic unit. Lock the seat by pressing the rear brake
and at least one of the two front brakes (refer to chapter 6).
4
Place the lumbar support cushion on the seat (refer to chapter 9).
5
Help the patient on to the seat, taking care to instruct the patient to
use exclusively the front part of the seat without attempting to get off
the side.
Place the patient's legs on the padded leg rest.
6
The front and rear shielding collars (optional) must be fixed to the
gantry and opened to the maximum limit, to avoid any type of
interference with insertion of the limb. Always leave the covers fitted
on the gantry.
fig. 14.1 - Position of the patient and seat during hand-wrist examinations
Содержание O-scan eXP
Страница 1: ......
Страница 2: ......
Страница 3: ...O scan User Manual 350004900 Rev 07 Esaote S p A Genoa...
Страница 9: ...350004900 Rev 07 7 10...
Страница 10: ...8 10...
Страница 12: ...10 10...
Страница 20: ...8 8...
Страница 32: ...8 8 Chapter 2...
Страница 51: ...350004900 Rev 07 19 48 Instructions for Safe Use fig 3 1 Top view...
Страница 52: ...20 48 Chapter 3 fig 3 2 Side view...
Страница 80: ...48 48 Chapter 3...
Страница 102: ...22 22 Chapter 4...
Страница 105: ...350004900 Rev 07 3 20 Technical Description fig 5 1 Equipment components...
Страница 122: ...20 20 Chapter 5...
Страница 126: ...4 20 Chapter 6 fig 6 3 The O scan Premium and O scan eXP patient seat with armrests fig 6 4 Armrest detail...
Страница 142: ...20 20 Chapter 6...
Страница 155: ...350004900 Rev 07 7 8 Coils fig 8 5 Spherical uniform phantom fig 8 6 Cylindrical uniform phantoms...
Страница 156: ...8 8 Chapter 8 fig 8 7 Geometric phantom fig 8 8 Centering support for the spherical phantom diameter 140 mm...
Страница 160: ...4 10 Chapter 9 fig 9 4 Lumbar support cushion fig 9 5 Positioning of the lumbar support cushion on the seat...
Страница 161: ...350004900 Rev 07 5 10 Soft cushion Set fig 9 6 Use of lumbar support cushion...
Страница 184: ...10 10 Chapter 11...
Страница 232: ...4 4 Chapter 17...
Страница 236: ...4 4 Chapter 18...