To prepare the analyzer for shipment or storage
1.
Shut off the process gas flow.
2.
Allow all residual gas to dissipate from the lines.
3.
Connect a purge supply, regulated to the specified sample supply
pressure, to the sample supply port.
4.
Confirm that any valves controlling the sample flow effluent to the
low pressure flare or atmospheric vent are open.
5.
Turn on the purge supply and purge the system to clear any residual
process gases.
6.
Turn off the purge supply.
7.
Allow all residual gas to dissipate from the lines.
8.
Close any valves controlling the sample flow effluent to the low
pressure flare or atmospheric vent.
9.
Disconnect power to the system.
10.
Disconnect all tubing and signal connections.
11.
Cap all inlets and outlets to prevent foreign material such as dust or
water from entering the system.
12.
Pack the equipment in the original packaging in which it was
shipped, if available. If the original packaging material is no longer
available, the equipment should be adequately secured (to prevent
excessive shock or vibration).
13.
If returning the analyzer to the factory, complete the
Decontamination Form provided by Hauser
attach to the outside of the shipping package as instructed before
shipping.
Storage
The packaged analyzer should be stored in a sheltered environment that is
temperature controlled between -20
C (-4
F) and 50
C (122
F), and should
not be exposed to direct sun, rain, snow, condensing humidity or corrosive
environments.
Disclaimers
Hauser accepts no responsibility for consequential damages arising
from the use of this equipment. Liability is limited to replacement and/or repair
of defective components.
This manual contains information protected by copyright. No part of this guide
may be photocopied or reproduced in any form without prior written consent
from Hauser.