SS500XP/SS2000XP Gas Analyzer
BA02183C
B
–
8
Hauser
5.
Open the door to the Analyzer. Refer to Figure B–3. In this view, the
old model pressure sensor is pictured.
6.
Remove the optical cable harness using a flat-head screwdriver.
7.
Disconnect the cell inlet using a 9/16 in. wrench.
8.
Disconnect the cell outlet using a 9/16 in. wrench.
9.
Disconnect the thermistor cable at the circular connector.
10.
Remove the pressure sensor cable from the circular connector inside
the enclosure.
For new model pressure sensors with quick-disconnects, detach the
pressure sensor cable from the pressure sensor at the connector
using a Phillips-head screwdriver. Do not remove the black connector
from the cable inside the enclosure.
Depending on the manner in which the analyzer is mounted,
removing the cell may not be necessary. If the cell is not removed
for pressure sensor replacement, skip to step 13.
Figure B–3
SS2000XP cabinet interior
CELL OUTLET
CELL INLET
MOUNTING
BRACKET
(behind cell)
OPTICAL HEAD
RIBBON CABLE
BACKPLANE
BOARD
PRESSURE
SENSOR/ CABLE
MOUNTING
BRACKET
(behind cell)