Hold the wrench on the flange stable and parallel to the
surface. Do not allow the 9/16 in. wrench to move.
b.
Turn the 7/8 in. wrench counterclockwise to loosen the pressure
sensor until it is able to be removed.
14.
Remove excess seal tape from the flange opening and threads and
check threads for galling. Refer to Figure B–6.
15.
Remove the new pressure sensor from the packaging. Retain the
black connector cap on the sensor — do not remove.
16.
Wrap stainless steel PTFE tape around the threads at the top of the
pressure sensor, beginning from the base of the threads to the top,
approximately three times taking care to avoid covering the top
opening. Refer to Figure B–7.
Threads showing signs of galling indicate a possible leak. Refer to
17 to arrange for repair.
Figure B–6
Removing excess seal tape from
flange
Figure B–7
Replacing seal tape