Instrument Problems
If the instrument does not appear to be hampered by gas leaks, contamination,
excessive sampling gas temperatures and pressures, or electrical noise, refer
to Table B–1 before contacting
Table B–1
Potential instrument problems and solutions
Symptom
Response
Non-Operation (at start up)
Is the power connected to both the analyzer and
power source? Is the switch on?
Non-Operation (after start up)
Is the power source good? (100 to 250 VAC @ 50
to 60 Hz, 9 to 16 VDC, 18 to 32 VDC).
Check fuse(s). If bad, replace with equivalent
amperage, slow-blow fuse.
Contact a factory sales representative for service
information.
Power Fail Error
Turn off the power to the unit and check the opti
-
cal head cables for a loose connection.
Do not
disconnect or reconnect any optical head
cables with the power connected.
Check the inlet and outlet tubes to see if they are
under any stress. Remove the connections to the
inlet and outlet tubes and see if the power goes
up. The existing tubing may need to be replaced
with stainless steel flexible tubing.
Possible mirror contamination issue. Contact a
factory sales representative for service informa
-
tion. If advised to do so, clean the mirrors by fol
-
lowing the instructions under
Possible alignment problem. Contact a factory
sales representative for service information.
Capture diagnostic data and send the file to
Hauser (refer to
“To read diagnostic
data with HyperTerminal”
in the Device
Parameters manual).
Verify Power Fail error. Refer to Device Parame
-
ters for instruction.