BA02183C
SS500XP/SS2000XP Gas Analyzer
Hauser
B
–
11
17.
Insert the new pressure sensor into the threaded flange keeping the
sensor parallel to the surface for proper fitting.
18.
Hand tighten the pressure sensor turning it counterclockwise into
the flange until no longer moving freely. Refer to Figure B–8.
19.
Using the 9/16 in. wrench to hold the flange in place, turn the sensor
clockwise with a 7/8 in. wrench until tight. Two or three threads on
the pressure sensor should still be visible.
20.
Remove the black connector from the pressure sensor and discard.
Make sure the black connector at the bottom of the pressure
sensor is facing up from the measurement cell, or forward
towards the enclosure door. Refer to Figure B–9.
Figure B–8
Replacing pressure sensor
Figure B–9
Newly installed pressure sensor
positioning