SS500XP/SS2000XP Gas Analyzer
Hauser
3
–
5
3.
For AC systems, pull ground, neutral and hot wires (#14 AWG
minimum) into the electronics enclosure.
For DC systems, pull ground, plus and minus wires.
4.
Strip back the jacket and/or insulation of the wires just enough to
connect to the power terminal block.
5.
For AC systems, attach the neutral and hot wires to the power
terminal block by
connecting the neutral wire to the terminal marked
“NEU” and the hot wire to the terminal marked “LINE” as shown in
For DC systems, connect the minus wire to the terminal marked “
”
and the positive wire to the terminal marked “+” as shown in Figure
6.
Connect the ground wire to the ground terminal marked .
7.
Close and tighten the electronics enclosure door.
An approved switch or circuit breaker rated for 15 amps should be
used and clearly marked as the disconnecting device for the
analyzer.
NEU
LINE
AC TERMINAL BLOCK
_
+
DC TERMINAL BLOCK
Figure 3–1
AC and DC connection terminal blocks in
electronics enclosure