BA02183C
SS500XP/SS2000XP Gas Analyzer
Hauser
B
–
3
Tools and supplies
•
Lens cleaning cloth (Cole Parmer
®
EW-33677-00 TEXWIPE
®
Alphawipe
®
Low-Particulate Clean Room Wipes or equivalent)
•
Reagent-grade isopropyl (ColeParmer
®
EW-88361-80 or equivalent)
•
Small drop dispenser bottle (Nalgene
®
2414 FEP Drop Dispenser
Bottle or equivalent)
•
Acetone-impenetrable gloves (North NOR CE412W Nitrile Chemsoft™
CE Cleanroom Gloves or equivalent)
•
Hemostat (Fisherbrand™ 13-812-24 Rochester-Pean Serrated
Forceps)
•
Bulb blower or dry compressed air/nitrogen
•
Torque wrench
•
Permanent marker
•
Flashlight
Determining the type of cell mirror
Measurement cells will come equipped with either a glass or stainless steel
mirror. The stainless steel mirrors are identified with either an “X” engraved on
the outside bottom of the mirror or a groove around the rim of the mirror. Glass
mirrors have no external markings. To determine the type of mirror being used
for the system cell:
1.
Feel at the bottom of the cell for the engraved “X” marking or
grooved rim. Refer to Figure B–1 below.
Always handle the optical assembly by the edge of the mount.
Never touch the coated surfaces of the mirror.
Figure B–1
Stainless steel mirror marking
MIRROR MARKED WITH ‘X’
MIRROR GROOVED RIM -
SIDE VIEW
Содержание SS2000XP
Страница 16: ...SS500XP SS2000XP Gas Analyzer BA02183C 1 12 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 36: ...SS500XP SS2000XP Gas Analyzer BA02183C A 8 Endress Hauser THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 59: ......